Made from the finest platinum and platinum-rhodium alloy, FLuke 5650 type R thermocouples and type S thermocouples cover a temperature range of 0 °C to 1450 °C with uncertainties as good as 0.15 °C over most of that range.
With four different models for each type, we have the type R thermocouple and type S thermocouple to fit your application. The measuring junction of both the 5649 and the 5650 is encased in a 0.25-inch (6.35 mm) alumina sheath that can be ordered in lengths of 20 or 25 inches (50.8 or 63.5 cm) to fit the specific requirements of your application. A reference, or “cold," junction may also be ordered. The reference junction uses a stainless steel sheath and is 8.25 inches long (21 cm) by 0.188 inches in diameter (4.8 mm). The thin diameter minimizes the immersion depth needed, but the extra length ensures you can get all the immersion you like.

Features
- Made from the finest platinum and platinum-rhodium alloy
- These type S thermocouples cover a temperature range of 0 °C to 1450 °C
- Special tin-plated
- Solid-copper connecting wires with ultra-low EMF properties

Major Applications
- Broadly used in many industrial and scientific applications
- All industrial markets including power generation, oil and gas, aerospace, semiconductors,
pharmaceutical, biotechnology, food processing and metals

Technical Specifications
-Range : 0 °C to 1450 °C
-Type : Platinum/10 % Rhodium vs. platinum (type S)
-Calibration : Fixed point calibration uncertainties (k=2): ± 0.15 °C up to 962 °C,
increasing linearly to ± 2.0 °C at 1450 °C
-Hot Junction
-Sheath Dimensions : 6.35 mm (0.25 in) diameter;
see Ordering Information for lengths

General Specifications
- Long-Term Stability ± 0.5 °C to 1100 °C
± 2.0 °C to 1450 °C (over one year depending on usage)
- Short-Term Stabilities : ± 0.2 °C to 1100 °C
± 0.6 °C to 1450 °C
- Immersion : At least 152 mm (6 in) recommended
- Protective Case Model 2609 case included